Flow meter accuracy
Kytola flow meter accuracy is stated as % of F.S. (Full Scale) or % of RD (Reading). Here you can find the information about flow meter’s accuracy and calculation formula for calculating inaccuracy.
Kytola flow meter accuracy is stated as % of F.S. (Full Scale) or % of RD (Reading). Here you can find the information about flow meter’s accuracy and calculation formula for calculating inaccuracy.
For example when flow meter full scale is 100 L/min and accuracy is stated as +/- 5 % F.S. it means +/- 5 L/min accuracy on the flow meter whole measurement range.
% of F.S. inaccuracy is calculated
Flow meter reading – Real flow *100 = F.S. inaccuracy
Flow meter full scale
For example when flow meter reading is 100 L/min and accuracy is stated as +/- 5 % RD it means +/- 5 L/min accuracy. Or at 50 L/min reading +/- 2,5 L/min accuracy.
% of RD. inaccuracy is calculated
Flow meter reading – Real flow *100 = RD inaccuracy