Country & language: USA / English
Distributors login

Flow meter accuracy

Kytola flow meter accuracy is stated as % of F.S. (Full Scale) or % of RD (Reading). Here you can find the information about flow meter’s accuracy and calculation formula for calculating inaccuracy.

% F.S. accuracy

For example when flow meter full scale is 100 L/min and accuracy is stated as +/- 5 % F.S. it means +/- 5 L/min accuracy on the flow meter whole measurement range.

% of F.S. inaccuracy is calculated

Flow meter reading – Real flow            *100 = F.S. inaccuracy
         Flow meter full scale

% RD accuracy

For example when flow meter reading is 100 L/min and accuracy is stated as +/- 5 % RD it means +/- 5 L/min accuracy. Or at 50 L/min reading +/- 2,5 L/min accuracy.

% of RD. inaccuracy is calculated

Flow meter reading – Real flow           *100 = RD inaccuracy
         Flow meter reading

Contact information

Let's get in touch

Kytola Instruments Inc
900 Old Roswell Lakes Parkway, Suite 120
Roswell, GA 30076

Hidden
Hidden
This field is for validation purposes and should be left unchanged.

Sales

flow-meter-accuracy

Sales (AR, LA, MS, NM, TX)

flow-meter-accuracy

President

flow-meter-accuracy

Where are you from?

Please select your country to browse in your own language and local units.